Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22421 |
| Brand Name: | Phadia ImmunoCAP Allergen c5 |
| Model: | |
| Affected: | Phadia ImmunoCAP™ Allergen c5, Ampicilloyl and ImmunoCAP™ Allergen c6, Amoxicilloyl, All lot numbers |
| Software version: | |
| Recalling Organisation: | Abacus ALS Limited 55b Richard Pearse Drive Airport Oaks Auckland 2022 |
| Contact Information: | 0800 222 170 |
| Manufacturer: | Phadia AB |
| Issue: | An issue has been found with the stability of ImmunoCAP Allergen c5, Ampicilloyl and ImmunoCAP Allergen c6, Amoxicilloyl. An internal monitoring real time stability study determined that the stability specifications were not met. The products have a nominal 12 months shelf life stated on the label. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 5/1/2018 |