Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:22394
Brand Name:Roche OptiView DAB IHC Detection Kit
Model:06396500001
Affected:LOts: Y11625, Y15571, Y19271,and Y24225
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:
Brand Name:Roche ultraView DAB IHC Detection Kit
Model:05269806001
Affected:Lots: Y09284, Y11687, Y11716, Y15384, Y17984, Y22153, Y18099, Y18069, Y19302 and Y22147
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:
Brand Name:OptiView Amplification Kit
Model:06396518001
Affected:Lots: Y15435, Y19322 and Y22447
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:
Brand Name:OptiView Amplification Kit (250 Test)
Model:06718663001
Affected:Lot: Y19318
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:
Brand Name:Hematoxylin II
Model:05277965001
Affected:Lots: Y10759, Y13938, Y17402, Y17403, Y21312 and Y22561
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:
Issue:Manufacturer has received increased customer complaints reporting leaking and sticking reagent dispensers. These reports are currently focused on horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.The manufacturer has identified the cause of the issue, and is working to correct it and additionally, has mandated specific requirements for slide controls for customers with affected product in inventory.
An expanded scope notification:
Two additional contributing factors were also identified:
1. Products intended for greater than 50 actuations
2. Products with emulsifiers
These two contributing factors, when combined, exacerbate the issue and its occurrence in the market.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/8/2018
Hide menus