Detail of Recall

Type of Product:Device
Medsafe Reference:22390
Brand Name:Simulator, invasive blood pressure
Model:
Affected:Model EVHRS
Software version:
Recalling Organisation:Edwards Lifesciences (New Zealand) Limited
171 Morrin Road
St John
Auckland 1072
Contact Information:021 313 234
Manufacturer:Edwards Lifesciences LLC
Issue:The manufacturer have received reports of Out of Range error messages for the Edwards ClearSight Heart Reference Sensor. These complaints have increased in the last six (6) months. An analysis of EVHRS Out of Range complaints led to the determination that the product is not meeting the current labelled shelf life (expiry date) of 730 days (2 years) from the date of manufacture
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/12/2017