Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22380 |
| Brand Name: | Medtronic Clip Gun Scalp Clip System |
| Model: | GUN CG8900 RANEY CLIP KIT 1PK model number CG8900, KIT CG8901 CLIP GUN 10 PK model number CG8901 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Issue: | Testing performed by the manufacturer showed that the outer sterile pouch of the product did not meet standards for package integrity. Specifically, the testing demonstrated that the outer pouch could be compromised and possibly go undetected by the user which could potentially lead to patient infections. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 19/12/2017 |