Detail of Recall

Type of Product:Device
Medsafe Reference:22374
Brand Name:Ventilators, continuous, life supporting
Model:CARESCAPE R860 ventilators with software version 10SP05
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Issue:There is a potential for the display processor to experience an unexpected failure. If this failure occurs, the device screen will temporarily go blank and will then display an Error Code "EC01 XX". The blank display results in loss of monitored data being displayed to the user.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/12/2017