Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:22361
Brand Name:VALITUDE X4 CRT-P U128
Model:
Affected:Model U128
Software version:
Recalling Organisation:Boston Scientific NZ Ltd
3 Diamond Street
Eden Terrace
Auckland 1021
Contact Information:0800 742 678
Manufacturer:Cardiac Pacemakers Inc
Brand Name:NG4 - RESONATE X4 CRT-D G447
Model:
Affected:Model G447
Software version:
Recalling Organisation:Boston Scientific NZ Ltd
3 Diamond Street
Eden Terrace
Auckland 1021
Contact Information:0800 742 678
Manufacturer:Cardiac Pacemakers Inc
Brand Name:NG4 - MOMENTUM CRT-D G124
Model:
Affected:Model G124, G125, G126, G128
Software version:
Recalling Organisation:Boston Scientific NZ Ltd
3 Diamond Street
Eden Terrace
Auckland 1021
Contact Information:0800 742 678
Manufacturer:Cardiac Pacemakers Inc
Brand Name:AUTOGEN CRT-D Model G172
Model:
Affected:Models G172, G173, G175, G177, G179
Software version:
Recalling Organisation:Boston Scientific NZ Ltd
3 Diamond Street
Eden Terrace
Auckland 1021
Contact Information:0800 742 678
Manufacturer:Boston Scientific Corporation
Brand Name:VISIONIST CRT-P U225
Model:
Affected:Models U225, U226, U228
Software version:
Recalling Organisation:Boston Scientific NZ Ltd
3 Diamond Street
Eden Terrace
Auckland 1021
Contact Information:0800 742 678
Manufacturer:Cardiac Pacemakers Inc
Brand Name:DYNAGEN CRT-D Model G151
Model:
Affected:Models G151, G156, G158
Software version:
Recalling Organisation:Boston Scientific NZ Ltd
3 Diamond Street
Eden Terrace
Auckland 1021
Contact Information:0800 742 678
Manufacturer:Boston Scientific Corporation
Issue:The manufacturer has included important programming information to prevent an unintended asynchronous biventricular (BiV) pacing behavior when tracking elevated atrial intrinsic rhythms in certain Cardiac Resynchronization Therapy (CRT) pacemakers (CRT-Ps) and defibrillators (CRT-Ds). Repeated detection of this unintended asynchronous BiV pacing behavior may result in the implanted device reverting to a permanent Safety Mode (Safety Core) status thus requiring early replacement.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Healthcare Professional
Date Commenced:15/12/2017