| Type of Product: | Device |
| Medsafe Reference: | 22355 |
| Brand Name: | Artis Q, Q.zen, zeego, zee systems |
| Model: | |
| Affected: | Artis systems, 5904433, 7008605, 7412807, 7413078, 7555357, 7555365, 7727717, 7728392, 10094135, 10094137, 10094139, 10094141, 10280959, 10848280, 10848281, 10848282, 10848283, 10848353, 10848355, 10848600, 10849000 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare GmbH |
| Issue: | The Artis table top carriage is mounted on top of the table base and fixed with a bolt. A snap ring is used to secure the bolt. In case of a missing or wrongly installed snap ring the bolt could slide out of its base, and the table top carriage will slide to the floor. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/12/2017 |