Detail of Recall

Type of Product:Device
Medsafe Reference:22354
Brand Name:ARCHITECT Cuvette Segment
Model:ARCHITECT c4000 Cuvette Segment, ARCHITECT c8000 Cuvette Segment, ARCHITECT c16000 Cuvette Segment
Affected:
Software version:
Recalling Organisation:Abbott Australasia Pty Ltd (NZ)
Abbott Laboratories NZ
4 Pacific Rise, Mt Wellington
Auckland 1060
Contact Information:61 2 93849717
Manufacturer:Abbott Laboratories
Issue:The base of the ARCHITECT Cuvette Segment may become detached under specific conditions. When a cuvette segment base is detached, cuvettes may be seated lower than the designed height. This may result in inadequate dispensing into specific cuvettes due to the sample probe being unable to make efficient contact with the cuvette bottom.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/12/2017