Detail of Recall

Type of Product:Device
Medsafe Reference:22332
Brand Name:iViewGT
Model:
Affected:All Digital Accelerators with iViewGT
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Limited (England)
Brand Name:XVI 4.5
Model:
Affected:All Digital Accelerators with XVI
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Limited (England)
Issue:The manufacturer has identified that if the microswitch in the PCB assembly, solenoid limit switch of the middle arm is not set up correctly, it is possible to get an uncontrolled extension of MV or kV detector arms. When the detector arm is fully extended there is no risk.

Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/11/2017