| Type of Product: | Device |
| Medsafe Reference: | 22329 |
| Brand Name: | Cleveland CT |
| Model: | |
| Affected: | Brilliance iCT, Brilliance iCT SP, and IQon Spectral CT systems |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems |
| Issue: | There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner's main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard coming from the gantry. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 29/11/2017 |