Detail of Recall

Type of Product:Device
Medsafe Reference:22314
Brand Name:CombiDiagnost R90
Model:
Affected:All CombiDiagnost
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:The manufacturer has become aware that in some CombiDiagnost deliveries the tube adapter-plate green/yellow protective earthing wire was removed
During GCF tubes assembly, spacers with different thickness are used to position pre-loc ring in respect to focal spot in specified distance.
There is a potential of an electrical shock if following factors occurs:
1. An electrical failure leads to an electrical potential on the adapter plate
2. The earth resistance based on mechanical connection is beyond limit
3. Person removes the tube cover. (not allowed for the customer)
4. Person touches the live wired part during exposure/fluoroscopy.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/11/2017