Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:22310
Brand Name:MITG - Battery pack (RMS)
Model:
Affected:Serial Numbers begin with 1201xxxxxx through 1712xxxxxx
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Covidien LLC
Issue:A functional battery installed in the PB980 ventilator automatically recharges when the battery depletes and the ventilator is connected to AC power. In the case of a battery with incorrect firmware, it may fail to fully charge. This situation does not impact the ventilator's operation when it is connected to AC power. However, if the ventilator is operated on battery power alone, this situation could limit the amount of time the ventilator is operational.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Sponsor Control
Date Commenced:23/11/2017