Detail of Recall

Type of Product:Device
Medsafe Reference:22304
Brand Name:Infusion pumps - Syringe (Swiss)
Model:
Affected:Product codes with prefix (all variants): 8001, 8002, 8003,8004, and 8005
Software version:
Recalling Organisation:CareFusion New Zealand 313 Limited
Unit 4 / 110 Mays Road
Onehunga
AUCKLAND 1061
Contact Information:0508 422 734
Manufacturer:CareFusion Switzerland 317 Sarl
Issue:The manufacturer has received a report regarding reoccurring PL3 alarms. A PL3 alarm is an internal malfunction error code, which will result in the pump stopping and alerting the user of an internal failure. If a pump displays a PL3 error message, the pump should be immediately removed from clinical use and sent for servicing by qualified service personnel.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/11/2017