Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22266 |
| Brand Name: | iQ200 Series Urine Microscopy Analyzer |
| Model: | |
| Affected: | iQ200 Series Urine Microscopy Analyzer with Barcode Reader Model NFT-2100 |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Inc |
| Issue: | The manufacturer has determined that the iQ200 Series Urine Microscopy Analyzer with Barcode Reader (Model NFT-2100), may intermittently exhibit the following: 1. Failure to read the urine sample dilution barcode label where it defaults to a dilution factor of 1:1 and does not apply the correct dilution factor. 2. Failure to read body fluid dilution barcode label where the rack is rejected and no results are generated. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 16/11/2017 |