Detail of Recall

Type of Product:Device
Medsafe Reference:22266
Brand Name:iQ200 Series Urine Microscopy Analyzer
Model:
Affected:iQ200 Series Urine Microscopy Analyzer with Barcode Reader Model NFT-2100
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:The manufacturer has determined that the iQ200 Series Urine Microscopy Analyzer with Barcode Reader (Model NFT-2100), may intermittently exhibit the following:
1. Failure to read the urine sample dilution barcode label where it defaults to a dilution factor of 1:1 and does not apply the correct dilution factor.
2. Failure to read body fluid dilution barcode label where the rack is rejected and no results are generated.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/11/2017