Detail of Recall

Type of Product:Device
Medsafe Reference:22246
Brand Name:Integrity
Model:
Affected:All Integrity™ R1.2 and Integrity™ R3.2 systems with a Precise Treatment Table for ATM license option
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Limited (England)
Issue:The Precise treatment table is equipped with sensors that achieve a high degree of reliability in the readout of the table position. In case of an undetected failure of one of these sensors, it is possible to get a positional error with the Precise Treatment Table following automatic table movement when using XVI and MOSAIQ.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/11/2017