Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22246 |
| Brand Name: | Integrity |
| Model: | |
| Affected: | All Integrity™ R1.2 and Integrity™ R3.2 systems with a Precise Treatment Table for ATM license option |
| Software version: | |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Elekta Limited (England) |
| Issue: | The Precise treatment table is equipped with sensors that achieve a high degree of reliability in the readout of the table position. In case of an undetected failure of one of these sensors, it is possible to get a positional error with the Precise Treatment Table following automatic table movement when using XVI and MOSAIQ. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 10/11/2017 |