Detail of Recall

Type of Product:Device
Medsafe Reference:22243
Brand Name:VITEK 2 Gram Positive- ID Card
Model:
Affected:VITEK® 2 GP ID Lots #2420192403, 2420198203
Software version:
Recalling Organisation:Medi'Ray New Zealand Ltd
53-55 Paul Matthews Road
Albany
Auckland 0632
Contact Information:0800 633 472
Manufacturer:Biomerieux SA
Issue:An internal investigation has confirmed, occurrences of atypical negative O129R reactions for Enterococcus casseliflavus ATCC® 700327™ in association with three(3) VITEK® 2 GP ID, causing failure of the Quality Control (QC) testing for the VITEK® 2 GP ID card. In accordance with the VITEK® 2 GP ID Instructions For Use, the expected O129R reaction for ATCC® 700327™ strain is positive. As a precaution, the manufacturer is also including in the scope of this notice all GP ID card lots manufactured with the same raw materials
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/11/2017