Detail of Recall

Type of Product:Device
Medsafe Reference:22096
Brand Name:Radiometer ABL800 series blood gas analyser with FLEXQ module
Model:
Affected:Sampler Registration receipt data, All serial numbers
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:It has come to the manufacturers attention that in certain situations, the registration receipt received after placing a sampler in the FLEXQ can be associated with a different sampler if the barcode quality on the receipt has been affected.
Update:
The manufacturer has now released the software upgrade Version6.18 which is the corrective action for the issue reported.
A checksum has been added to the barcode on the registration receipt to make the barcode more robust.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/1/2018