Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22096 |
| Brand Name: | Radiometer ABL800 series blood gas analyser with FLEXQ module |
| Model: | |
| Affected: | Sampler Registration receipt data, All serial numbers |
| Software version: | |
| Recalling Organisation: | Radiometer Pacific Suite 1 1/30 Benedicts Street Eden Terrace Auckland 1010 |
| Contact Information: | 0800 723 722 |
| Manufacturer: | Radiometer Medical ApS |
| Issue: | It has come to the manufacturers attention that in certain situations, the registration receipt received after placing a sampler in the FLEXQ can be associated with a different sampler if the barcode quality on the receipt has been affected. Update: The manufacturer has now released the software upgrade Version6.18 which is the corrective action for the issue reported. A checksum has been added to the barcode on the registration receipt to make the barcode more robust. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 9/1/2018 |