Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22085 |
| Brand Name: | Fisher and Paykel Healthcare Catheter Mount |
| Model: | RT021 |
| Affected: | From Lot 2100089358 to Lot 2100098462 and from Lot 2100190360 to Lot 2100241254 |
| Software version: | |
| Recalling Organisation: | Fisher & Paykel Healthcare Ltd 15 Maurice Paykel Place East Tamaki AUCKLAND 2013 |
| Contact Information: | 0800 503 553 |
| Manufacturer: | Fisher & Paykel Healthcare Ltd |
| Issue: | The manufacturer is initiating a recall due to an increased rate of split tubing cuffs in catheter mounts manufactured on specific dates. A split tubing cuff on the catheter mount may potentially cause a leak which could lead to patient hypoxemia if undetected. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 17/10/2017 |