Detail of Recall

Type of Product:Device
Medsafe Reference:22074
Brand Name:Datex GE Carestation Anaesthesia Work Stations
Model:630, 650 and 650c
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Issue:It has been identified that an incomplete seal can exist between the CO2 absorber and the breathing circuit CO2 bypass port assembly of the Carestation 600 Series systems. An incomplete seal can allow rebreathing of patient gases that have bypassed the CO2 absorbent material and could result in unintended elevated levels of inspired CO2 (FiCO2), which could lead to hypercarbia. There have been no injuries reported as a result of this issue.
Manufacturer is releasing revised parts that minimize the likelihood of incomplete gas flow through the CO2 absorbent canister. GEHC will correct all affected products by installing updated parts.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/10/2017