Detail of Recall

Type of Product:Device
Medsafe Reference:22073
Brand Name:Heartware Ventricular Assist Device (HVAD) Controllers
Model:
Affected:Serial numbers lower than CON300000
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:HeartWare Inc
Brand Name:Heartware Ventricular Assist Device (HVAD) DC Adapters
Model:1435
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:HeartWare Inc
Issue:As described in previous Hazard Alerts from April 2015 and April 2016, there was the potential for the following safety issues associated with the current HeartWare HVAD System Controller, including:
1. Worn alignment guides, potentially resulting in damaged connector pins.
2. Internal double disconnect (no power) alarm battery failure, which could prevent the controller from Sounding an alarm in the event of a complete interruption of power.
3. Loose power and data connectors, which could allow fluid ingress
With the introduction of the updated HVAD Controller, also referred to as Controller 2.0, Medtronic is initiating removal procedures for previous generation HeartWare HVAD Controllers and all HeartWare DC Adapters, which are incompatible with the new HVAD Controller. The removal of these HVAD Controllers and DC Adapters will occur concurrently with the introduction of the new HVAD Controller.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Healthcare Professional
Date Commenced:10/10/2017