Detail of Recall

Type of Product:Device
Medsafe Reference:22071
Brand Name:Smiths Medical 50 ml CADD® Medication Cassette
Model:21-7001-24
Affected:Lot Nos: multiple (expiration date range March 2017 and March 2022)
Software version:
Recalling Organisation:Smiths Medical Australasia Pty Ltd
c/o KPMG, KPMG Centre
18 Viaduct Harbour Avenue
Auckland
Contact Information:0800 444 200
Manufacturer:Smiths Medical MD Inc
Brand Name:Smiths Medical 100 ml CADD Administrations Set
Model:21-7002-24
Affected:Lot Nos: multiple (expiration date range March 2017 and March 2022)
Software version:
Recalling Organisation:Smiths Medical Australasia Pty Ltd
c/o KPMG, KPMG Centre
18 Viaduct Harbour Avenue
Auckland
Contact Information:0800 444 200
Manufacturer:Smiths Medical MD Inc
Brand Name:Smiths Medical 100ml CADD Medication cassette, Yellow
Model:21-7100-24
Affected:Lot Nos; multiple (expiration date range March 2017 and March 2022)
Software version:
Recalling Organisation:Smiths Medical Australasia Pty Ltd
c/o KPMG, KPMG Centre
18 Viaduct Harbour Avenue
Auckland
Contact Information:0800 444 200
Manufacturer:Smiths Medical MD Inc
Issue:The manufacturer has become aware that certain Non Flow-Stop CADD Medication Cassette reservoirs may have been manufactured with an incorrect pressure plate. Normally the pressure plate has a tunnel for the pumping tube, however on the affected models there is an arch on the pressure plate and pumping tube is pushed up.
Under delivery of medication may result from the tubing becoming partially or completely occluded when the affected cassette is attached to the pump.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/10/2017