Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22071 |
| Brand Name: | Smiths Medical 50 ml CADD® Medication Cassette |
| Model: | 21-7001-24 |
| Affected: | Lot Nos: multiple (expiration date range March 2017 and March 2022) |
| Software version: | |
| Recalling Organisation: | Smiths Medical Australasia Pty Ltd c/o KPMG, KPMG Centre 18 Viaduct Harbour Avenue Auckland |
| Contact Information: | 0800 444 200 |
| Manufacturer: | Smiths Medical MD Inc |
| Brand Name: | Smiths Medical 100 ml CADD Administrations Set |
| Model: | 21-7002-24 |
| Affected: | Lot Nos: multiple (expiration date range March 2017 and March 2022) |
| Software version: | |
| Recalling Organisation: | Smiths Medical Australasia Pty Ltd c/o KPMG, KPMG Centre 18 Viaduct Harbour Avenue Auckland |
| Contact Information: | 0800 444 200 |
| Manufacturer: | Smiths Medical MD Inc |
| Brand Name: | Smiths Medical 100ml CADD Medication cassette, Yellow |
| Model: | 21-7100-24 |
| Affected: | Lot Nos; multiple (expiration date range March 2017 and March 2022) |
| Software version: | |
| Recalling Organisation: | Smiths Medical Australasia Pty Ltd c/o KPMG, KPMG Centre 18 Viaduct Harbour Avenue Auckland |
| Contact Information: | 0800 444 200 |
| Manufacturer: | Smiths Medical MD Inc |
| Issue: | The manufacturer has become aware that certain Non Flow-Stop CADD Medication Cassette reservoirs may have been manufactured with an incorrect pressure plate. Normally the pressure plate has a tunnel for the pumping tube, however on the affected models there is an arch on the pressure plate and pumping tube is pushed up. Under delivery of medication may result from the tubing becoming partially or completely occluded when the affected cassette is attached to the pump. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/10/2017 |