Detail of Recall

Type of Product:Device
Medsafe Reference:22052
Brand Name:Reagent Fluorocell Plt
Model:
Affected: Lots 6051 – 6090. 
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Sysmex Corporation
Issue:some of the reagent Fluorocell Plt within the affected lots are susceptible to external environmental factors and tend to decrease their dye concentration over time. This deterioration resulted in incorrect Plt-F results with a high discrepancy between the Plt-I value and a falsely low Plt F value.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/10/2017