Detail of Recall

Type of Product:Device
Medsafe Reference:22031
Brand Name:Medtronic O-arm O2 Surgical Imaging System
Model:
Affected:model number BI-700-02000
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Navigation Inc
Issue:During servicing of certain internal components of the O-armTM O2, it was identified that the high-power cable connector may come in contact with surrounding metal (toggle switch or corner of battery tray) during the disconnecting/reconnecting process. If this occurs it could result in an electrical short and a shock to the field service technician.The manufacturer will be installing a service kit on the O-armTM O2 surgical imaging system to further enhance the safety of field service personnel.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/9/2017