Detail of Recall

Type of Product:Device
Medsafe Reference:22026
Brand Name:Artis Q, Q.zen, zeego, zee systems
Model:
Affected:Artis systems with wireless footswitch
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Brand Name:Artis one
Model:
Affected:Artis systems with wireless footswitch
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Shenzhen Magnetic Resonance Ltd
Issue: The wireless foot switch can fail due to impacts by external factors, such as electrostatic discharge that exceeds a certain intensity. If the wireless foot switch fails, it will no longer be possible to use it to release radiation. It will still be possible to trigger exposures if a wired foot switch or hand switch is present. Fluoroscopy with the hand switch is not possible.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/9/2017