Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 22024 |
| Brand Name: | Infusion - Swiss - Connectors Adaptors |
| Model: | |
| Affected: | 11 gauge 9391-0200 Lot numbers 1470303001, 1470304601, 1470306301,1470311301, 1470314801, 1570317701,1570317702, 1570317704, 1570322001,1470306602 |
| Software version: | |
| Recalling Organisation: | CareFusion New Zealand 313 Limited Unit 4 / 110 Mays Road Onehunga AUCKLAND 1061 |
| Contact Information: | 0508 422 734 |
| Manufacturer: | CareFusion Switzerland 317 Sarl |
| Issue: | The investigation results indicate that the pinholes in the blister pouch is related to the secondary packaging process where the blister packs are placed in dispensers and in shelf boxes. The device has sharp points and during the packing of the carton, accidental pressure was applied, causing the blister pouches to press against each other. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 26/9/2017 |