Detail of Recall

Type of Product:Device
Medsafe Reference:22021
Brand Name:Brilliance iCT/iCT SP
Model:
Affected:CT systems running software version 4.1.6 are potentially affected
Software version:4.1.6
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:Upon completion of scan procedure during Bolus tracking on a Brilliance iCT with software version 4.1.6, no images were generated. This leads to unavailability of raw data for offline reconstruction.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/9/2017