Detail of Recall

Type of Product:Device
Medsafe Reference:22011
Brand Name: Coated Implant_Delta Extend
Model:
Affected:Modular Centered Epiphysis Size 1 HA, Modular Eccentric Epiphysis Size 1 Left HA, Modular Eccentric Epiphysis Size 1 Right HA, Modular Centered Epiphysis Size 2 HA, Modular Eccentric Epiphysis Size 2 Left HA, Modular Eccentric Epiphysis Size 2 Right HA
Software version:
Recalling Organisation:Johnson & Johnson (NZ) Limited
13a Gabador Place
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:DePuy France SA
Issue:The manufacturer has determined that lots produced in a specific machine were manufactured in a manner in which the post's diameter is out-of-specification. Further investigation determined that the issue is related to the circularity of the cylinder. The circularity issue may cause the epiphysis to interfere with the stem and the two implants (epiphysis and stem) cannot be assembled.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Healthcare Professional
Date Commenced:21/9/2017