Type of Product:Device
Medsafe Reference:21995
Brand Name:Portable Detector Holder
Model:
Affected:clamp issue
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:The manufacturer has become aware that if the lock mechanism of the movable clamp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below the detector) and additional force is applied. This may happen when the detector is rotated clockwise from landscape to portrait position.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/9/2017