Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21995 |
| Brand Name: | Portable Detector Holder |
| Model: | |
| Affected: | clamp issue |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems DMC GmbH |
| Issue: | The manufacturer has become aware that if the lock mechanism of the movable clamp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below the detector) and additional force is applied. This may happen when the detector is rotated clockwise from landscape to portrait position. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 19/9/2017 |