Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21975 |
| Brand Name: | Lysercell WDF |
| Model: | |
| Affected: | Affected Lots: A6027, A6028, A6029, A6030, A6031, A6032, A6033, A6034, A6035, A6037, A6039, A7001, A7002, A7004 |
| Software version: | |
| Recalling Organisation: | Roche Diagnostics (NZ) Ltd 15 Rakino Way Mt Wellington AUCKLAND 1644 |
| Contact Information: | 0800 363 622 |
| Manufacturer: | Sysmex Corporation |
| Issue: | It has been noticed that there is an increased frequency of false positive flags in the WDF channel. Upon investigation, the manufacturer determined there was a possibility of XN series instruments generating this problem due to certain lots of Lysercell WDF. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 14/9/2017 |