Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:21967
Brand Name:Radiometer AQT90 FLEX Immunoassay analyzer
Model:
Affected:Ambient storage conditions for on-board test cartridges
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:the manufacturer has recently become aware that the internal temperature in the AQT90 FLEX Analyzer under worst case conditions may slightly exceed acceptable storage conditions for AQT90 FLEX on-board test cartridges.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/9/2017