Detail of Recall

Type of Product:Device
Medsafe Reference:21959
Brand Name:Artis Q, Q.zen, zeego, zee systems
Model:
Affected:Lot number – 10094137, 10848281, 10848282, 10848280, 10848283, 10094135, 10094139, 10848355, 10280959, 10848353
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Issue:In the application "DSA Roadmap" the Artis system's intended behavior is to reject the stored Vesselmap from previously acquired DSA if the patient table was moved during the Roadmap workflow. In rare cases the system does not reject the DSA Vesselmap and uses it for the further Roadmap workflow steps. Depending on the amount of table movement which was applied during the Roadmap workflow, the DSA Vesselmap might be overlaid as vessel tree at a position which does not fulfill the requirements for accuracy.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/9/2017