Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21930 |
| Brand Name: | Low rise bed - Ningbo |
| Model: | |
| Affected: | Hi Lo Bed PHCBED1 |
| Software version: | |
| Recalling Organisation: | Pride Mobility Products New Zealand Limited 391 Neilson Street Penrose Auckland 1061 |
| Contact Information: | |
| Manufacturer: | Ningbo Kangmailong Medical Apparatus Co. Ltd |
| Issue: | The side rails supplied with the beds do not comply with the standard IEC 60601-2-52:2009: Medical electrical equipment - Part 2-52: Particular requirements for the basic safety and essential performance of medical beds and there is a risk of patient harm, including but not limited to patient entrapment |
| Recall action type: | Correction |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/9/2017 |