Detail of Recall

Type of Product:Device
Medsafe Reference:21922
Brand Name:Slit lamps HS010, SL 120, SL 130, SL 115 'Classic'
Model:
Affected:Serial Numbers: 970712XXXX and 970714XXXX
Software version:
Recalling Organisation:Carl Zeiss (NZ) Ltd
22 Fremlin Place
Avondale
Auckland 1026
Contact Information:09 838 5626
Manufacturer:Carl Zeiss Ophthalmic Systems AG
Issue:The manufacturer is aware that in certain devices there is an increased probability of 1) failure of the power supply ON/OFF switch, and 2) surface corrosion of the mirror which changes the direction of the light projection. Therefore the power supply ON/OFF switch and the coating of the mirror has been modified to improve their reliability.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/9/2017