Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21922 |
| Brand Name: | Slit lamps HS010, SL 120, SL 130, SL 115 'Classic' |
| Model: | |
| Affected: | Serial Numbers: 970712XXXX and 970714XXXX |
| Software version: | |
| Recalling Organisation: | Carl Zeiss (NZ) Ltd 22 Fremlin Place Avondale Auckland 1026 |
| Contact Information: | 09 838 5626 |
| Manufacturer: | Carl Zeiss Ophthalmic Systems AG |
| Issue: | The manufacturer is aware that in certain devices there is an increased probability of 1) failure of the power supply ON/OFF switch, and 2) surface corrosion of the mirror which changes the direction of the light projection. Therefore the power supply ON/OFF switch and the coating of the mirror has been modified to improve their reliability. |
| Recall action type: | Correction |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 6/9/2017 |