Detail of Recall

Type of Product:Device
Medsafe Reference:21919
Brand Name:Hamilton/47244/Ventilators
Model:
Affected:Hamilton C2 and Hamilton C3
Software version:
Recalling Organisation:InterMed Medical Ltd
71 Apollo Drive
Albany
AUCKLAND 1311
Contact Information:0800 333 444
Manufacturer:Hamilton Medical AG
Issue:The manufacturer has identified an issue relating to the battery charge of the HAMILTON-C2/C3 batteries after a usage of two and a half years. After two and a half years, the battery fuel gauge may indicate a higher battery charge as to what is available. The technical failure 444001 low battery high priority alarms are triggered, alerting the operator both acoustically and visually before entering ambient mode.
Recall action type:Correction
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/9/2017