Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21882 |
| Brand Name: | Aquios CL Flow Cytometry System |
| Model: | |
| Affected: | Product number (PN B30166) |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Inc |
| Issue: | The manufacturer has learned that an erroneous configuration setting during servicing may be possible on the AQUIOS CL instruments. If the incorrect configuration occurs, the probe may aspirate samples only from a single tube and will not aspirate samples from the remaining tubes in the cassette. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 25/8/2017 |