Detail of Recall

Type of Product:Device
Medsafe Reference:21866
Brand Name:AQUIOS CL Flow Cytometry System
Model:
Affected:Aquios CL flow Cytometry system
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:The manufacturer has learned that an erroneous configuration setting during servicing may be possible on the AQUIOS CL instruments. If the incorrect configuration occurs, the probe may aspirate samples only from a single tube and will not aspirate samples from the remaining tubes in the cassette.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/8/2017