Detail of Recall

Type of Product:Device
Medsafe Reference:21823
Brand Name:Abbott RealTime HIV-1 Qualitative
Model:
Affected:List 04N66-90
Software version:
Recalling Organisation:Abbott Australasia Pty Ltd (NZ)
Abbott Laboratories NZ
4 Pacific Rise, Mt Wellington
Auckland 1060
Contact Information:61 2 93849717
Manufacturer:Abbott Laboratories
Issue:For plasma samples, RealTime HIV-1 Qualitative may exhibit false Not Detected results for HIV samples between 110 Copies/mL (Limit of Detection) and 336 Copies/mL. For Dried Blood Spot (DBS) samples, Abbott RealTime HIV-1 Qualitative performs in accordance with the detection rate within the RealTime HIV-1 Qualitative package insert for DBS sample types. No action is required for DBS samples.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/8/2017