Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21822 |
| Brand Name: | Immunoassay Premium Quality Controls |
| Model: | |
| Affected: | Catalogue Number IA2638 All lots |
| Software version: | |
| Recalling Organisation: | Abacus ALS Limited 55b Richard Pearse Drive Airport Oaks Auckland 2022 |
| Contact Information: | 0800 222 170 |
| Manufacturer: | Randox Laboratories Ltd |
| Issue: | The manufacturer has confirmed that ACTH in the Immunoassay Premium quality control does not meet the quoted reconstituted stability claim in the product IFU. The manufacturer recommends that ACTH is to be assayed immediately after the 30-minute reconstitution procedure. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/8/2017 |