Detail of Recall

Type of Product:Device
Medsafe Reference:21816
Brand Name:Tourniquet Systems
Model:
Affected:updating an on-device label on ATS2200 and ATS4000 units
Software version:
Recalling Organisation:Zimmer Biomet New Zealand Company
210 Khyber Pass Road
Grafton
Auckland
Contact Information:09 925 5200
Manufacturer:Zimmer Surgical, Inc.
Issue:In a typical OR setting, the manufacturer originally recommended a minimum separation distance of 3.8 meters (m) between RF communication equipment to mitigate the potential for electromagnetic interference and to achieve optimal performance of the tourniquet machines. RF equipment consists of, but is not limited to, electrosurgical equipment, including any consoles, wiring, and pencil units. This distance recommendation originates from IEC 60601-1-2. As a result of customer feedback, additional internal testing was performed to see if this distance could be shortened, and it was determined that 1.0 meter provided adequate distance to avoid interference.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/8/2017