| Type of Product: | Device |
| Medsafe Reference: | 21800 |
| Brand Name: | Cleveland CT |
| Model: | |
| Affected: | Brilliance iCT SP (728311), Brilliance iCT (728306) Running software version: 4.1.6 |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems |
| Issue: | During a bolus tracking procedure: If the operator sets the ROI on the locator slice and then presses GO to initiate the tracker scans, the following message may display: Scan Parameters Exceeds System resources. If the message is displayed, the system will fail to proceed with the tracker scan. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 9/8/2017 |