Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21791 |
| Brand Name: | ProBP 2400 Digital Blood Pressure |
| Model: | |
| Affected: | (21)07150001 - (21)07150620, (21)12150001 - (21)12150500 |
| Software version: | |
| Recalling Organisation: | Universal Specialities Ltd trading as USL Medical 494 Rosebank Road Avondale AUCKLAND 1026 |
| Contact Information: | 09 829 0595 |
| Manufacturer: | Welch Allyn Inc |
| Issue: | As a result of a defective electrical component in the battery charging circuit, when the ProBP 2400 is connected to the external power supply/battery charger, a potential over-voltage battery charging condition may occur. This potential defect resides with the device, not the battery. The over-voltage condition can result in high NiMH battery temperatures that, in some cases, can reach levels sufficient to cause melting of the plastic (ABS) battery door and other adjacent plastic and foam device components. |
| Recall action type: | Correction |
| Recall action: | Product to be exchanged |
| Level of recall: | Wholesale |
| Date Commenced: | 14/8/2017 |