Detail of Recall

Type of Product:Device
Medsafe Reference:21791
Brand Name:ProBP 2400 Digital Blood Pressure
Model:
Affected:(21)07150001 - (21)07150620, (21)12150001 - (21)12150500
Software version:
Recalling Organisation:Universal Specialities Ltd trading as USL Medical
494 Rosebank Road
Avondale
AUCKLAND 1026
Contact Information:09 829 0595
Manufacturer:Welch Allyn Inc
Issue:As a result of a defective electrical component in the battery charging circuit, when the ProBP 2400 is connected to the external power supply/battery charger, a potential over-voltage battery charging condition may occur. This potential defect resides with the device, not the battery.
The over-voltage condition can result in high NiMH battery temperatures that, in some cases, can reach levels sufficient to cause melting of the plastic (ABS) battery door and other adjacent plastic and foam device components.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Wholesale
Date Commenced:14/8/2017