Detail of Recall

Type of Product:Device
Medsafe Reference:21737
Brand Name:VITEK MS INSTRUMENT
Model:
Affected:V2 or V3 update : ref 413654,415706, 417104, 418884 and 411032, V3: ref 420260 and 421661
Software version:
Recalling Organisation:Medi'Ray New Zealand Ltd
53-55 Paul Matthews Road
Albany
Auckland 0632
Contact Information:0800 633 472
Manufacturer:BioMerieux Inc
Issue:It has been identified that the VITEK® MS system could give, in specific conditions, an incorrect identification result if the tested species is not included in the VITEK® MS knowledge base (KB). This is a system limitation with all MALDI-TOF databases which is currently not described in the VITEK® MS documentation.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/7/2017