Type of Product:Device
Medsafe Reference:21735
Brand Name:Pro BP Handheld BP Device
Model:
Affected:Serial Numbers (21)12150001 - (21)12150500
Software version:
Recalling Organisation:Ebos Group Ltd
108 Wrights Road
Addington
Christchurch
Contact Information:0800 18 17 16
Manufacturer:Welch Allyn Inc
Issue:The manufacturer has been made aware of an issue which could potentially affect the ProBP 2400 Digital Blood Pressure. As a result of a defective electrical component in the battery charging circuit, when the ProBP 2400 is connected to the external power supply/battery charger, a potential over-voltage battery charging condition may occur. This potential defect resides with the device, not the battery. The over-voltage condition can result in high NiMH battery temperatures that, in some cases, can reach levels sufficient to cause melting of the plastic (ABS) battery door and other adjacent plastic and foam device components.
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/7/2017