Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21730 |
| Brand Name: | Balloon pump |
| Model: | |
| Affected: | CS100 and CS300 Intraortic Balloon Pumps, also applies to any System 98 or System 98XT IABP which was converted to a CS100i or CS300 IABP. |
| Software version: | |
| Recalling Organisation: | Maquet Australia Pty Ltd (NZ) 12G Andromeda Cres East Tamaki Auckland 2013 |
| Contact Information: | 021 874 882 |
| Manufacturer: | Datascope Corporation |
| Issue: | The manufacturer has received complaints involving the CS100/CS100i and CS300 IABPs regarding the following issues: " False blood detection alarm, and " The ingress of fluids into the IABP affecting various electronic circuit boards. Either issue could potentially prevent initiation or continuation of therapy. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 27/7/2017 |