Detail of Recall

Type of Product:Device
Medsafe Reference:21726
Brand Name:Infant Warmers
Model:
Affected:Affected model numbers: Model 2001 IWS (International), 3000 IWS, 3050 IWS, 3100 IWS, 3150 IWS, 3300 IWS, 3500 IWS, 4000 IWS, 4300 IWS, 4400 IWS, 5000 IWS.
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Ohmeda Medical Inc
Issue:The manufacturer has recently become aware of a potential safety issue related to loose screws in the Heater Head of certain Infant Warmer System devices. Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/7/2017