Detail of Recall

Type of Product:Device
Medsafe Reference:21703
Brand Name:Philips HeartStart MRx or XL+ monitor/defibrillator
Model:
Affected:sync cables, models M1783A and M5526A, and ECG out cables, model 989803195641.
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:When a monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable or ECG out cable, the following can occur if the monitor/defibrillator experiences interference from electrical fast transients (EFTs):
" On the HeartStart MRx and HeartStart XL, EFT noise may be misinterpreted as an R-wave.
" On the HeartStart XL+, EFT noise can disable ECG monitoring and potentially interrupt demand mode pacing.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/7/2017