Detail of Recall

Type of Product:Device
Medsafe Reference:21695
Brand Name:Patient Reference Frame ( Unclassified)
Model:
Affected:9734715 Spinous Process Clamp, Tall, 9734716 Spinous Process Clamp, Short, 9734723 Double Spinous Process Clamp, Tall, 9734724 Double Spinous Process Clamp, Short, 9734715K Spinous Process Clamp, Tall, 9734716K Spinous Process Clamp, Short and 9734723K
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Navigation Inc
Issue:The manufacturer has become aware of instances where Spine Clamps have been damaged when forced open beyond their intended limits during use and will subsequently not open once attached to a patient's spinous process. When the Spine Clamp is forced open beyond its intended limits, a component (captive washer) may break off. The washer could then be inadvertently left behind in a patient's body if the breakage occurs during the procedure.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/7/2017