Detail of Recall

Type of Product:Device
Medsafe Reference:21692
Brand Name:Microgenics Application Sheets
Model:
Affected:Various Microgenics Appliction Sheets, All Lots
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Microgenics Corporation
Issue:During an internal assessment the manufacturer found that Application Sheets provided by Microgenics for expanded use of some Assays are inadequate. Specifically, the appropriate validation data for the expanded use of these Assays, as indicated in the Instructions for Use, when not used on Manufacturer approved analyzers, is inadequate. In addition Microgenics did not clearly state in these Application Sheets that the parameters provided are only to be used as starting points and that each Assay setting needs to be validated individually by the user.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/7/2017