Type of Product:Device
Medsafe Reference:21690
Brand Name:Smith & Nephew Osteotome
Model:
Affected:Thin and Radial Osteotome Blades (Model and Batch specific)
Software version:
Recalling Organisation:Smith & Nephew Surgical Ltd
Unit A 36 Hillside road
Wairau Valley
AUCKLAND 0760
Contact Information:0800 657 799
Manufacturer:Smith & Nephew Inc
Issue:The manufacturer has become aware that the affected devices are labelled with expiration dates based on a 10 year packaging shelf-life where the actual validated shelf life of the outer pouch packaging is 5 years. This could potentially result in the outside of the inner tray packaging being non-sterile, potentially contaminating the sterile field
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/7/2017