Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21687 |
| Brand Name: | Sysmex |
| Model: | |
| Affected: | CellaVision DI-60 |
| Software version: | |
| Recalling Organisation: | Roche Diagnostics (NZ) Ltd 15 Rakino Way Mt Wellington AUCKLAND 1644 |
| Contact Information: | 0800 363 622 |
| Manufacturer: | Roche Diagnostics GmbH |
| Issue: | It has been reported that a possible mix up of results could occur during 'user modification' of a barcode misread in the following circumstances: 1. DI-60 encountered a problem reading the barcode when processing a slide (Slide 1) and a picture of the barcode for Slide 1 is taken by the DI-60. 2. When this occurs the DI-60 provides Slide 1 with a temporary order ID based on the date and time (ie ERRYYYYMMDDhhmmss) and the image of the barcode picture (taken in step 1) is stored in the database and displayed in the order data dialogue against this ERR ID. 3. Following this, a slide (Slide 2) from the slide maker/stainer (SP-10) which has had a barcode misread occurring on the SP-10 and has been. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 14/7/2017 |